    DESCRIPTION

    The Midmark M11 Steam Sterilizer is designed to be safe, reliable and easy to use. With one of the largest chambers of any tabletop autoclave on the market, the M11 can optimize your supply of available sterile instruments by reducing the time and effort required for sterilization.

    Reliable   

    Steam-flush pressure pulse air removal system ensures proper steam sterilization.

    Easy to Use

    Pre-programmed sterilization cycles and intuitive LCD display with prompts ensure quick, confident operation.

    Safe

    Real-time alerts notify users promptly of any critical system errors

    Overall length w/plug: 18.1 in (50.0 cm)

    Overall width: 17.8" (45.2 cm)

    Overall height w/printer: 17.8" (45.2 cm)

    Minimum countertop area: 21" D x 17.8" W (53.3 cm x 45.2 cm) 

    Chamber: 11" diameter (28 cm) 18" depth (45.7 cm) 6.5 gal usable volume (24.6 liters)

    Trays: Two large – 9" W x 15" L x 1.1" D (38 cm x 2.9 cm) Two small – 6.6" W x 15" L x 1.1" D (16.8 cm x 38 cm x 2.9 cm)

    Unit weight: 99 lb (44.9 kg)

    Shipping weight: 131 lb (59.4 kg)

    Water reservoir capacity: 1.4 gal (5.3 liters)

    Electrical requirements: 115 VAC, 50/60 Hz, 15 A single phase 115 VAC has 1,425 W tubular immersion heater 230 VAC, 50/60 Hz, 10 A single phase 230 VAC has 1,500 W tubular immersion heater

    Installation: Use on a level, water-resistant material such as laminate, stainless-steel or stone surface. Allow 2.0" (5.08 cm) on sides and rear, and 5" (12.7 cm) above to allow space to fill reservoir and change the printer paper roll.

    Safety standards: UL listed to US and Canada standards UL 61010-1, 2nd edition IEC 61010-2-040, 1st edition CAN/CSA C22.2 No. 61010-1, 2nd edition CSA C22.2 No. 61010-2-040-07, Part 2-040, 1st edition Meets the requirements of ASME Boiler and Pressure Vessel Code. A separate (dedicated) circuit is recommended for this sterilizer.

    Certifications: IEC 60601-1-2 Edition 4.0 2014-02 CISPR II: 2009, A1:2010 Class A, Group 1 IEC 61000-3-2:2014, Part 3-2 IEC 61000-3-3:2014, Part 3-3 ANSI/AAMI ST55:2010(R) 2014 ANSI/AAMI/IEC 62304

     