No Products in the Cart
The tube revolves freely around the joint, allow it to reach practically any position, including the vertical. RXDC - HyperSphere technology also features an automatic touch-sensitive device for simple, efficient locking/release of the X-ray head tube so it can be repositioned effortlessly between one exposure and the next. Ergonomic zones on the sides of the head provide a firm grip for effective positioning.
Simple and immediate: Wireless
The wireless remote controller lets the user control the device (by communicating with the X-ray tube) while enjoying full freedom of movement.
Access to exposure programmes is provided via two simple settings. The large display shows the sequential exposure monitor and the patient exposure dose; moreover, the controller has a wireless X-ray snapshot button. Wireless device control allows fast, easy installation: no fixed control panels are required, thus providing greater freedom when positioning the X-ray unit.
MAXIMUM QUALITY
With a tiny focal spot of 0.4 mm (at 30 cm), RXDC - HyperSphere technology produces sharp images under any condition.
The tube head is now even more powerful as it operates at 70 kV, 8 mA. RXDC - HyperSphere technology gives your surgery the precision and quality of cuttingedge know-how.
MULTI-MODE
Maximum flexibility to meet your diagnostic needs. Automatic parameter modulation ensures exposure power and time are always selected according to the patient’s build and the specific region of investigation.
SEQUENCED EXPOSURE
The dynamic service cycle allows uninterrupted use of the RXDC, as in the case of systematic examinations, and real-time monitoring of tube head temperature on the large wireless controller display.
COMPLETE EQUIPMENT SET
RXDC - HyperSphere technology features a 30 cm cone with a circular and rectangular collimator. The rectangular collimator is ideal for reducing the X-ray dose administered to the patient as it limits the irradiated body area to the effective capture capacity of the sensor.
Technical Specifications
TECHNICAL DATA | ||
Generator | Constant potential, microprocessor-controlled | |
Working frequency | 145-230 KHz (typically 175 KHz) | |
Focal spot | 0,4 mm (IEC 336) | |
Anode current | 4 / 8 mA | |
Voltage at X-ray tube | 60 / 65 / 70 kV (*) | |
Exposure time | 0.020 – 1.000 seconds, R’10 and R’20 scale | |
Source-skin distance | 20 e 30 cm | |
Irradiated field | 35 x 45 mm rectangular, Ø 55 mm and Ø 60 mm round | |
Additional collimators | 31 x 41 mm and 22 x 35 mm, for size 2 and size 1 sensors | |
Total filtration | 2.0 mm Al @ 70kV | |
Power supply | 50/60 Hz, 115-120Vac ±10% o 230-240Vac ±10% | |
Duty Cycle | continuous operation with self-adjustment up to 1s/80s total | |
Stability | Automatic lock/release, with touch-sensitive activation (HyperSphere technology) | |
Arms | Available in 3 lengths: 40 cm - 60 cm - 90 cm | |
Max. arm extension | 230 cm, from wall | |
Certification | CE 0051, FDA approved |
(*) values depend on the country where the product is marketed.